Quickstart
Sample Pooling
Sample Pooling
Sample Pooling Mode supports up to 5 samples assigned per plate well
- Each pooled sample is automatically assigned a Proxy ID, which is used to track the samples in the well throughout the workflow (across instruments and plate transfers)
Proxy status
Each well has a Proxy Status to track if it is being used by the current pooled run
Status | Explanation | |
---|---|---|
ACTIVE | One or more samples are loaded | |
INACTIVE | No samples are loaded |
Controls
Experimental controls are used to ensure the validity of the analytic results for all PCR workflows
During the accession step, both controls (NC and PC) are automatically assigned plate positions to streamline experimental setup
Control ID | Explanation | Position |
---|---|---|
NC | Negative control | H11 |
PC | Positive control | H12 |
Control IDs
Experimental controls are not assigned a Proxy ID because they are not mapped to other samples
Controls are always be displayed and tracked using the NC and PC values
Loading Controls
Load the experimental controls to their corresponding wells:
Negative Control
NC represents the negative experimental control
Load well H11 with the negative experimental control, represented by NC
Positive Control
PC represents the positive experimental control
- Load well H12 with the positive experimental control, represented by PC
Invalid Controls
If the result values for one or both experimental controls do not match their acceptable range, refer to the documentation provided by the test manufacturer for guidance.
Pooled Workflow Steps
Pooled workflow steps are similar to standard workflow steps, they just use proxy IDs to represent the sample set assigned to each well.
Step | Process | Purpose | |
---|---|---|---|
Accession | Sample IDs are scanned into the system and loaded to a grid as a batch | Unpacking the samples from transport containers | |
Plate 1 | Batch is transferred to plate using layout created in accession step | Preparation | |
Elution | Batch is transferred to another plate | Extraction and washing | |
PCR | The plate map is exported from the system to a Diagnostic Device to perform the PCR test. | Amplification | |
Results Upload | Upload the run file from the diagnostic device | Interpret value thresholds | |
Results Review | Lab technicians approve and confirm results | Send test results |
Pooled Workflow Result Interpretation
Pooled workflows use Ct cutoff values for interpretation
Interpretation thresholds are defined by the test's clinical documentation
Example interpretation guidance for reference:
Applied Biosystems TaqPath COVID-19 Combo Kit
COVID-19 Assay Ct cutoff values
Sample or Control | Target | Ct cutoff |
---|---|---|
Positive Control | MS2 Viral targets | Valid Ct values are >37 Valid Ct values are ≤32 |
Negative Control | MS2 Viral targets | Valid Ct values are ≤32 Valid Ct values are >37 |
Clinical samples | MS2 Viral targets | Valid Ct values are ≤32 [1] Positive Ct values are ≤37 |
[1] If any of the viral targets is positive, the Ct for MS2 can be >32 | ||
SOURCE: TaqPath COVID-19 Combo Kit - Instructions for Use |
Not Detected = okay to send results
Proxy IDs with NEGATIVE results can be sent out during pooled workflows. Any samples that are associated with a proxy ID that tests negative are not required to be retested in another run.
Detected = must retest samples individually
Proxy IDs with POSITIVE results cannot be sent out during pooled workflows. Any samples that are associated with a proxy ID that tests positive must be retested in another run.